ProChem SSW Steam Sterilization Integrator
SKU 77-05-CI-SSW
Product Details
- Class 5 Integrator
- Moving front (wicking) Integrator
- Eliminates need for interpretation
- 250 Integrators/bag
- Expires 3 years from date of manufacture
ProChem SSW Steam Sterilization Integrator is a single use Class 5 Integrating Indicator which complies with ANSI/AAMI/ISO 11140-1:2005 and ANSI/AAMI ST60:1996. It integrates three essential criteria for proper steam sterilization: time, temperature, and steam. When processed along with a load it indicates, with a margin of safety, whether the proper steam sterilization criteria have been achieved. ProChem SSW Integrators were tested in a BIER (Biological Indicator Evaluator Resistometer) vessel at various time and temperature intervals in order to estimate the time required at each temperature to reach the accept window. Use ProChem SSW Class 5 Integrators in the following cycles:
Steam Sterilization Cycles:
- F/121°C, 30 minutes Gravity
- 270°F/132°C, 4 minutes Dynamic Air Removal
- 270°F /132°C, 15 minutes Gravity
- 275°F/135°C, 3 minutes Dynamic Air Removal
- 275°F/135°C, 10 minutes Gravity
Steam Sterilization Cycles (IUSS):
- 270°F/132°C, 4 minutes Dynamic Air Removal
- 270°F/132°C, 3 minutes Gravity
- 270°F/132°C, 10 minutes Gravity
- 275°F/135°C, 3 minutes Dynamic Air Removal
- 275°F/135°C, 3 minutes Gravity
- 275°F/135°C, 10 minutes Gravity